| MSOX6004A X-Series specification overview | |
| Bandwidth (-3dB) | 1 GHz | 
| Calculated rise time (10 to 90%) | ≤ 350 ps | 
| Input channels | 4 + 16 | 
| Maximum sample rate | 20 GSa/s half channels, 10 GSa/s full channels | 
| Maximum memory depth | 4 Mpts half channels, 2 Mpts all channels | 
| Display size and type | 12.1-inch capacitive multi-touch/gesture-enabled display | 
| Waveform update rate | > 450,000 waveforms per second | 
| Vertical system analog channels | |
| Hardware bandwidth limits | 1 MΩ: 20 MHz, 200 MHz (selectable per channel) 50 Ω: 20 MHz, 200 MHz, 1.5 GHz, 3 GHz (selectable per channel)  | 
		
| Input coupling | AC, DC | 
| Input impedance | Selectable: 1 MΩ ± 1% (14 pF), 50 Ω ± 3% | 
| Input sensitivity range | 1 mV/div to 5 V/div at 1 MΩ (200 MHz bandwidth limit at ≤ 2 mV/div) 1 mV/div to 1 V/div at 50 Ω  | 
		
| Vertical resolution | 8 bits (measurement resolution is 12 bits with averaging) | 
| Maximum input voltage | 1 MΩ: 30 Vrms or ± 40 Vmax (DC + Vpeak); Probing technology allows testing of higher voltages. The included N2894A 10:1 probe supports 300 Vrms or ± 400 Vmax (DC + Vpeak) 50 Ω: ± 5Vpk max  | 
		
| DC vertical gain accuracy | ± 3 div: 1.5% of full scale (typical)± 4 div: 2.5% of full scale (warranted) | 
| DC vertical offset accuracy | ± 0.1 div ± 2 mV ± 1% of offset setting (valid for an offset of ≤ ± 12 divisions) | 
| Channel-to-channel isolation | ≥ 100:1 (DC to 1 GHz), ≥ 30:1 (> 1 GHz) | 
| Vertical system digital channels | |
| Digital input channels | 16 digital (D0 to D15. pod 1: D7 ~ D0, Pod 2: D15 ~ D8) | 
| Thresholds | Threshold per pod | 
| Threshold selections | TTL (+1.4 V), 5 V CMOS (+2.5 V), ECL (-1.3 V), user-defined (selectable by pod) | 
| User-defined threshold range | ± 8.0 V in 10 mV steps | 
| Maximum input voltage | ± 40 V peak | 
| Threshold accuracy | ± (100 mV + 3% of threshold setting) | 
| Maximum input dynamic range | ± 10 V about threshold | 
| Minimum voltage swing | 500 mVpp | 
| Input impedance | 100 kΩ ± 2% at probe tip | 
| Input capacitance | ~ 8 pF | 
| Vertical resolution | 1 bit | 
| Display characteristics | |
| Display | 12.1-inch capacitive multi-touch/gesture enabled color TFT LCD | 
| Display mode | Zone/zoom/annotation mode and waveform placement mode | 
| Resolution | 800 (H) x 480 (V) pixel format (screen area) | 
| Graticules | 8 vertical divisions by 10 horizontal divisions with intensity controls | 
| Format | YT, XY, and Roll | 
| Maximum waveform update rate | > 135,000 wfm/s (real time) > 450,000 wfm/s (real time max update rate)  | 
		
| Persistence | Off, infinite, variable persistence (100 ms to 60 s) | 
| Intensity gradation | 256 intensity levels | 
| General and environmental characteristics | |
| Power line consumption | Max 200 W | 
| Power voltage range | 100 to 120 V, 50/60/400 Hz 100 to 240 V, 50/60 Hz  | 
		
| Environmental rating | 0 to +50 °C; 3962 m Max Maximum Relative Humidity (non-condensing): 80%RH up to 40 °C From 40 °C to 50 °C, the maximum % Relative Humidity follows the line of constant dew point  | 
		
| Safety | ANSI/UL Std. No. 61010-1:2012; CAN/CSA-C22.2 No. 61010-1-12 | 
| Vibration | Meets IEC60068-2-6 and MIL-PRF-28800; class 3 random | 
| Shock | Meets IEC 60068-2-27 and MIL-PRF-28800; class 3 random Operating 30 g, ½ sine. 11 ms duration, 3 shocks/axis along major axis, total of 18 shocks  | 
		
| Dimensions (W x H x D) | 425 mm W x 288 mm H x 148 mm D (16.7 x 11.3 x 5.8)in.  | 
		
| Weight | Net: 6.8 kg (15 lbs.), Shipping: 11.3 kg (25 lbs.) | 