Keysight E4407B ESA Spectrum Analyzer

No stress rentals! We help you to choose the equipment that you need, we expedite shipment, and follow up.

Keysight E4407B ESA Spectrum Analyzer
  • Application

  • The E4407B provides laboratory grade performance up to 26.5 GHz. It is rugged enough for use in the field and accurate enough for applications such as aerospace and defense.
  • Key features

  • Frequency range: 9 kHz to 26.5 GHz
  • Frequency response 13.2 to 26.5 GHz ± 2 dB
  • Amplitude accuracy ± 0.34 dB, ± 0.13 dB typical
  • Average continuous power +30 dBm (1 W)
  • Peak pulse power +50 dBm (100 W)

Frequency specifications
Frequency range 9 kHz to 26.5 GHz
Measurement speed
Local measurement and display update rate 33 ms, (30/s)
Remote measurement and GPIB transfer rate 33 ms, (30/s)
Marker peak search 300 ms
Center frequency tune and transfer (RF) < 90 ms
Center frequency tune and transfer (μW) 350 ms
Offset from carrier signal. Typical in italic.
10 kHz –98 dBc/Hz, –101 dBc/Hz, (Option 1D5)
100 kHz –118 dBc/Hz, –122 dBc/Hz
1 MHz –125 dBc/Hz, –127 dBc/Hz
–133 dBc/Hz, –136 dBc/Hz (with option 120)
10 MHz –131 dBc/Hz, –136 dBc/Hz
–137 dBc/Hz, –141 dBc/Hz (with option 120)

Amplitude specifications
Mechanical input attenuator range 0 to 75 dB in 5 dB steps
Average continuous power +30 dBm (1 W)
Peak pulse power +50 dBm (100 W)
DC voltage 0 Vdc
50 Vdc
Displayed average noise level (dBm) (input terminated, 0 dB attenuation, sample detector) specifications
Typical values shown in italic
Frequency: 1 to 10 MHz -137 dBm, -139 dBm
Frequency: 10 to 500 MHz
Frequency: 500 MHz to 1 GHz
-135 dBm, -139 dBm
Frequency: 1 to 1.5 GHz
Frequency: 1.5 to 2 GHz
-135 dBm, -140 dBm
Frequency: 2 to 3 GHz
Frequency: 3 to 6 GHz
-131 dBm, -138 dBm
Frequency: 6 to 12 GHz -130 dBm, -137 dBm
Frequency: 12 to 22 GHz -126 dBm, -134 dBm
Frequency: 22 to 26.5 GHz -125 dBm, -132 dBm
Displayed average noise level (dBm) with RF preamplifier
Typical values shown in italic
Frequency: 1 to 10 MHz -155 dBm
Frequency: 10 MHz to 1 GHz -151 dBm, -157 dBm
Frequency: 1 to 2 GHz -151 dBm, -155 dBm
Frequency: 2 to 3 GHz -149 dBm, -152 dBm
Resolution bandwidth switching uncertainty
1 Hz, 3 Hz RBW ± 0.3 dB (Option 1D5)
10 Hz, 30 Hz RBW ± 0.3 dB
100 Hz, 300 Hz RBW ± 0.3 dB
1 kHz to 1.5 MHz RBW
1.5 to 3 MHz RBW
± 0.3 dB
5 MHz RBW ± 0.6 dB

General specifications
Operating temperature 0 to +55 °C
Storage temperature -40 to +75 °C
EMI compatibility Conducted and radiated interference is in compliance with CISPR Pub. 11/1990 Group 1 Class A. Conducted and radiated interference is in compliance with CISPR Pub. 11/1990 Group 1 Class Ba (Option 060)
Military specifications Type tested to the environmental specifi cations of MIL-PRF-28800F Class 3
AC operation 195 to 250 Vrms, 47 to 66 Hz
Power consumption < 300 W
DC operation 12 to 20 Vdc, < 200 W power consumption
Internal data storage 200 traces or states/8.0 MB
Dimensions Width to outside of instrument handle: 416 mm (16.4 in)
Overall height: 222 mm (8.75 in)
Depth from front frame to rear frame: 409 mm (16.1 in)
Weight Instrument: 17.1 kg (37.7 lbs)
Shipping: 31.9 kg (70.3 lbs)
For detailed specifications please see: Keysight E4407B Data Sheet