No stress rentals! We help you to choose the equipment that you need, we expedite shipment, and follow up.
| Frequency range | 9 kHz to 6.7 GHz | 
| Measurement speed | |
| Local measurement and display update rate | 33 ms, (30/s) | 
| Remote measurement and GPIB transfer rate | 33 ms, (30/s) | 
| Marker peak search | 300 ms | 
| Center frequency tune and transfer (RF) | < 90 ms | 
| Center frequency tune and transfer (μW) | 350 ms | 
| Offset from carrier signal. Typical in italic. | |
| 10 kHz | –98 dBc/Hz, –101 dBc/Hz, (Option 1D5) | 
| 100 kHz | –118 dBc/Hz, –122 dBc/Hz | 
| 1 MHz | –125 dBc/Hz, –127 dBc/Hz –133 dBc/Hz, –136 dBc/Hz (with option 120) | 
| 10 MHz | –131 dBc/Hz, –136 dBc/Hz –137 dBc/Hz, –141 dBc/Hz (with option 120) | 
| Mechanical input attenuator range | 0 to 75 dB in 5 dB steps | 
| Average continuous power | +30 dBm (1 W) | 
| Peak pulse power | +50 dBm (100 W) | 
| DC voltage | 0 Vdc 50 Vdc | 
| Displayed average noise level (dBm) (input terminated, 0 dB attenuation, sample detector) specifications Typical values shown in italic | |
| Frequency: 1 to 10 MHz | -137 dBm, -139 dBm | 
| Frequency: 10 to 500 MHz Frequency: 500 MHz to 1 GHz | -135 dBm, -139 dBm | 
| Frequency: 1 to 1.5 GHz Frequency: 1.5 to 2 GHz | -135 dBm, -140 dBm | 
| Frequency: 2 to 3 GHz Frequency: 3 to 6 GHz | -131 dBm, -138 dBm | 
| Displayed average noise level (dBm) with RF preamplifier Typical values shown in italic | |
| Frequency: 1 to 10 MHz | -155 dBm | 
| Frequency: 10 MHz to 1 GHz | -151 dBm, -157 dBm | 
| Frequency: 1 to 2 GHz | -151 dBm, -155 dBm | 
| Frequency: 2 to 3 GHz | -149 dBm, -152 dBm | 
| Resolution bandwidth switching uncertainty | |
| 1 Hz, 3 Hz RBW | ± 0.3 dB (Option 1D5) | 
| 10 Hz, 30 Hz RBW | ± 0.3 dB | 
| 100 Hz, 300 Hz RBW | ± 0.3 dB | 
| 1 kHz to 1.5 MHz RBW 1.5 to 3 MHz RBW | ± 0.3 dB | 
| 5 MHz RBW | ± 0.6 dB | 
| Operating temperature | 0 to +55 °C | 
| Storage temperature | -40 to +75 °C | 
| EMI compatibility | Conducted and radiated interference is in compliance with CISPR Pub. 11/1990 Group 1 Class A. Conducted and radiated interference is in compliance with CISPR Pub. 11/1990 Group 1 Class Ba (Option 060) | 
| Military specifications | Type tested to the environmental specifi cations of MIL-PRF-28800F Class 3 | 
| AC operation | 195 to 250 Vrms, 47 to 66 Hz Power consumption < 300 W | 
| DC operation | 12 to 20 Vdc, < 200 W power consumption | 
| Internal data storage | 200 traces or states/8.0 MB | 
| Dimensions | Width to outside of instrument handle: 416 mm (16.4 in) Overall height: 222 mm (8.75 in) Depth from front frame to rear frame: 409 mm (16.1 in) | 
| Weight | Instrument: 17.1 kg (37.7 lbs) Shipping: 31.9 kg (70.3 lbs) | 
| For detailed specifications please see: Keysight E4404B Data Sheet | |